G T KMB HDL D LDL D T HbA1c T T U T A A C C C H M P P R LFA - LICOPROTEÍNAS SLOTEST URBIDIMÉTRICO OLESTEROL IRETO OLESTEROL IRETO ERRITINA URBIDIMÉTRICA EMOGLOBINA ICROALBUMINÚRIA CRTEST URBIDIMÉTRICO CRTEST LTRASENSÍVEL EUMATEST URBIDIMÉTRICO 1 F ÁCIDA URBIDIMÉTRICO COBAS MIRA PLUS Manual de Automação REVISÃO: 22 (03/2013)

Upload: doankhue

Post on 07-Mar-2018

267 views

Category:

Documents


23 download

TRANSCRIPT

Page 1: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

GTKMBHDL DLDL D

THbA1c

TTUT

AA

CCC

HM

PP

R

LFA - LICOPROTEÍNAS SLOTEST URBIDIMÉTRICO

OLESTEROL IRETOOLESTEROL IRETO

ERRITINA URBIDIMÉTRICAEMOGLOBINA

ICROALBUMINÚRIA CRTEST URBIDIMÉTRICOCRTEST LTRASENSÍVEL

EUMATEST URBIDIMÉTRICO

1

F

ÁCIDA

URBIDIMÉTRICO

COBAS MIRA PLUS

Manual de AutomaçãoREVISÃO: 22 (03/2013)

Page 2: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

ÍNDICE

SAC: 0800 644 6433 ww w.doles.com .br

Manual de Automação - COBAS MIRA PLUS

ALBUMINA ...............................................................................................................01

ALFA 1-GLICOPROTEÍNAS ÁCIDA (LANÇAMENTO) .....................................................02

ALT/TGP-UV .............................................................................................................03

AMILASE CNPG .......................................................................................................04

ASLOTEST TURBIDIMÉTRICO (LANÇAMENTO) ..........................................................05

AST/TGO-UV ............................................................................................................06

CÁLCIO ARSENAZO ................................................................................................07

CKMB (LANÇAMENTO) ..................................................................................................08

CK-NAC ...................................................................................................................09

CLORETO COLORIMÉTRICO ................................................................................10

COLESTEROL ENZ. LÍQUIDO ................................................................................11

COLESTEROL HDL ................................................................................................12

COLESTEROL HDL DIRETO (LANÇAMENTO) .............................................................13

COLESTEROL LDL DIRETO (LANÇAMENTO) ..............................................................14

CREATININA ............................................................................................................15

DHL-UV ....................................................................................................................16

FERRITINA TURBIDIMÉTRICO (LANÇAMENTO) ..........................................................17

FOSFATASE ALCALINA CINÉTICA ...........................................................................18

FOSFATO-UV ...........................................................................................................19

GAMA GT- CINÉTICA ...............................................................................................20

GLICOSE ENZ. LÍQUIDA.........................................................................................21

HEMOGLOBINA HbA1c (LANÇAMENTO) ......................................................................22

MICROALBUMINÚRIA TURBIDIMÉTRICA (LANÇAMENTO) .........................................23

MICROPROTE pirogalol .........................................................................................24

PCRTEST TURBIDIMÉTRICO (LANÇAMENTO) ...........................................................25

PCRTEST ULTRASENSÍVEL (LANÇAMENTO) .............................................................26

PROTEÍNAS TOTAIS ...............................................................................................27

REUMATEST TURBIDIMÉTRICO (LANÇAMENTO) .......................................................28

TRIGLICÉRIDES ENZ. LÍQUIDO ............................................................................29

URATO ENZ. LÍQUIDO ............................................................................................30

UREIA -UV ...............................................................................................................31

REVISÃO: 22 (03/2013)

Page 3: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

SAC: 0800 644 6433 ww w.doles.com .br 01

ALBUMINAManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : R-SCALIBRATION MODE : SLOPE AVGREAGENT BLANK : REAG/DILCLEANER : BEFOREWAVELENGTH : 600nmDECIMAL POSITION : 2UNIT : g/dL

ANALYSISPOST DIL FACTOR : 2

CONC.FACTOR : NOSAMPLE CYCLE : 1

VOLUME : 2µLDILUENT NAME : H20

VOLUME : 20µLREAGENT CYCLE : 1

VOLUME : 300µL

CALCULATIONSAMPLE LIMIT : 0.1500

POINT : T1REAC. DIRECTION : INCREASE

CHECK : ONCONVERS. FACTOR : 1.000

OFF SET : 0.000TEST RANGE LOW : 0

HIGH : 8NORM. RANGE LOW : * *

HIGH : * *

NUMBER OF STEPS : 1CALC. STEP A : ENDPOINTREADINGS FIRST : CB LAST: 4

CALIBRATIONCALIB. INTERVAL : * *

TIME : * *

BLANKREAG. RANGE LOW : * *

HIGH : * *

BLANK RANGE LOW : * *

HIGH : * *

STANDARD POS : *

STD 1 :STD 2 :STD 3 :

REPLICATE : * *

DEVIATION : * *

CONTROLCS1 POS : * *

CS2 POS : * *

CS3 POS : * *

∗∗∗∗∗ Utilizar o valor assinalado do Calibrador Multiparâmetro.∗∗ ∗∗ ∗∗ ∗∗ ∗∗ Parâmetro a ser definido pelo usuário.

ALBUMINARef. A : 333 determinaçõesRef. B : 666 determinações

CONTROLE DE QUALIDADEO uso de soro controle de referência deve ser uma prática rotineira do laboratório. Recomenda-se utilizar um soro controle com valor na faixade normalidade (soro controle N - Doles) e outro soro controle de valor elevado (soro controle P - Doles).

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS.

Page 4: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

SAC: 0800 644 6433 ww w.doles.com .br 02

ALFA 1 GLICOPROTEÍNAS ÁCIDAManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : D-R-S-SR1CALIBRATION MODE : LOGIT/LOG 5REAGENT BLANK : REAG/DILCLEANER : NOWAVELENGTH : 340nmDECIMAL POSITION : 0UNIT : mg/dL

ANALYSISDILUENT NAME : H20

FACTOR : NOTIME : NOSTD : MAIN INDIRECT

MAIN STD : *POS : **

∗∗∗∗∗ Utilizar o valor assinalado do Calibrador que acompanha o kit.

FACT. STD 1 : 1.00 2 : 2.003 : 2.00 4 : 2.005 : 2.00 6 : 2.007 : NO 8 : NO

POST DIL FACTOR : 2CONC.FACTOR : NO

SAMPLE CYCLE : 2VOLUME : 2µL

DILUENT NAME : H20VOLUME : 5µL

REAGENT CYCLE : 1VOLUME : 240µL

START R1 CYCLE : 1VOLUME : 60µL

DILUENT NAME : H20VOLUME : 5µL

CALCULATIONSAMPLE LIMIT : NOREAC. DIRECTION : INCREASE

CHECK : OFFANTIGEN EXCESS : NO

CONVERS. FACTOR : 1.000OFF SET : 0.000

TEST RANGE LOW : OFFHIGH : OFF

NORM. RANGE LOW : **HIGH : **

NUMBER OF STEPS : 1CALC. STEP A : END POINTREADINGS FIRST : T1 LAST : 12REACTION LIMIT : NO

CALIBRATIONCALIB. INTERVAL : * *

TIME : * *

BLANKREAG. RANGE LOW : NO

HIGH : NOBLANK RANGE LOW : NO

HIGH : NOSTANDARDS : O aparelho irá mostrar as

concentrações correspondentespara cada diluição do calibrador

REPLICATE : * *

DEVIATION : * *

CONTROLCS1 POS : * *

CS2 POS : * *

CS3 POS : * *

∗∗ ∗∗ ∗∗ ∗∗ ∗∗ Parâmetro a ser definido pelo usuário.

ALFA 1 GLICOPROTEÍNAS ÁCIDA

Ref.A : 83 determinações

Ref.B :166 determinações

CONTROLE DE QUALIDADE

O uso de soro controle de referência deve ser uma prática

rotineira do laboratório. Recomenda-se utilizar um soro

controle com valor na faixa de normalidade e outro controle

de valor elevado.

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS

PROPRIETÁRIOS.

Page 5: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

SAC: 0800 644 6433 ww w.doles.com .br 03

ALT/TGP UVManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : R-SCALIBRATION MODE : SLOPE AVGREAGENT BLANK : REAG/DILCLEANER : NOWAVELENGTH : 340nmDECIMAL POSITION : 0UNIT : UI/L

ANALYSISPOST DIL FACTOR : 2

CONC. FACTOR : NOSAMPLE CYCLE : 1

VOLUME : 16µLDILUENT NAME : H20

VOLUME : 30µLREAGENT CYCLE : 1

VOLUME : 145µL

CALCULATIONSAMPLE LIMIT : 0.700

POINT : T1REAC. DIRECTION : DECREASE

CHECK : ONCONVERS. FACTOR : 1.000

OFF SET : 0.000TEST RANGE LOW : 0

HIGH : 350NORM. RANGE LOW : * *

HIGH : * *

NUMBER OF STEPS : 1CALC. STEP A : KINSEARCHREADINGS FIRST : 3 LAST: 14REACTION LIMIT : 0.2

POINT : T1CALIBRATIONCALIB. INTERVAL : * *

TIME : * *

BLANKREAG. RANGE LOW : NO

HIGH : NOBLANK RANGE LOW : -0.0020

HIGH : 0.0020

STANDARD POS : *

STD 1 :STD 2 :STD 3 :

REPLICATE : * *

DEVIATION : * *

CONTROLCS1 POS : * *

CS2 POS : * *

CS3 POS : * *

∗∗∗∗∗ Utilizar o valor assinalado do Calibrador Multiparâmetro.∗∗ ∗∗ ∗∗ ∗∗ ∗∗ Parâmetro a ser definido pelo usuário.

ALT/TGP UVRef. A : 344 determinaçõesRef. B : 689 determinações

PREPARO DO REAGENTE DE USOVide Instruções de uso do kit.

CONTROLE DE QUALIDADEO uso de soro controle de referência deve ser uma prática rotineira do laboratório. Recomenda-se utilizar um soro controle com valor na faixade normalidade (soro controle N - Doles) e outro soro controle de valor elevado (soro controle P - Doles).

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS.

Page 6: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

04

AMILASE CNPGManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : R-SCALIBRATION MODE : SLOPE AVGREAGENT BLANK : REAG/DILCLEANER : NOWAVELENGTH : 405nmDECIMAL POSITION : 0UNIT : UI/L

ANALYSISPOST DIL FACTOR : 2 CONC. FACTOR : NOSAMPLE CYCLE : 1

VOLUME : 3.0µLDILUENT NAME : H20

VOLUME : 10.0µLREAGENT CYCLE : 1

VOLUME : 140µL

CALCULATIONSAMPLE LIMIT : 0.300

POINT : T1REAC. DIRECTION : INCREASE

CHECK : ONCONVERS. FACTOR : 1.000

OFF SET : 0.000TEST RANGE LOW : 0

HIGH : 2000NORM. RANGE LOW : * *

HIGH : * *

NUMBER OF STEPS : 1CALC. STEP A : KINETICREADINGS FIRST : 1 LAST : 4REACTION LIMIT : 0.3100

POINT : T1CALIBRATIONCALIB. I NTERVAL : * *

TIME : * *

BLANKREAG. RANGE LOW : NO

HIGH : NOBLANK RANGE LOW : NO

HIGH : NO

STANDARD POS : *

STD 1 :STD 2 :STD 3 :

REPLICATE : * *

DEVIATION : * *

CONTROLCS1 POS : * *

CS2 POS : * *

CS3 POS : * *

∗∗∗∗∗ Utilizar o valor assinalado do Calibrador Multiparâmetro.∗∗ ∗∗ ∗∗ ∗∗ ∗∗ Parâmetro a ser definido pelo usuário.

AMILASE CNPGRef. A : 214 determinaçõesRef. B : 428 determinações

CONTROLE DE QUALIDADEO uso de soro controle de referência deve ser uma prática rotineira do laboratório. Recomenda-se utilizar um soro controle com valor na faixade normalidade (soro controle N - Doles) e outro soro controle de valor elevado (soro controle P - Doles).

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS.

SAC: 0800 644 6433 ww w.doles.com .br

Page 7: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

SAC: 0800 644 6433 ww w.doles.com .br 05

ASLOTEST TURBIDIMÉTRICOManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : R-SCALIBRATION MODE : SLOPE AVGREAGENT BLANK : REAG/DILCLEANER : NOWAVELENGTH : 550nmDECIMAL POSITION : 0UNIT : U.I/mL

ANALYSISPOST DIL FACTOR : 2

CONC.FACTOR : NOSAMPLE CYCLE : 1

VOLUME : 3µLDILUENT NAME : H20

VOLUME : 10µLREAGENT CYCLE : 1

VOLUME : 300µL

CALCULATIONSAMPLE LIMIT : NOREAC. DIRECTION : INCREASE

CHECK : ONCONVERS. FACTOR : 1.000

OFF SET : 0.000TEST RANGE LOW : 0 U.I/mL

HIGH : 800 U.I/mLNORM. RANGE LOW : 0 U.I/mL

HIGH : 200 U.I/mLNUMBER OF STEPS : 1CALC. STEP A : ENDPOINTREADINGS FIRST : 2 LAST: 7

CALIBRATIONCALIB. INTERVAL : **

BLANKREAG. RANGE LOW : NO

HIGH : NOBLANK RANGE LOW : NO

HIGH : NO

STANDARD POS : **STD 1 : *STD 2 : NOSTD 3 : NO

REPLICATE : **DEVIATION : **

CONTROLCS1 POS : **CS2 POS : **CS3 POS : **

∗∗∗∗∗ Utilizar o valor assinalado do Calibrador que acompanha o kit.∗∗ ∗∗ ∗∗ ∗∗ ∗∗ Parâmetro a ser definido pelo usuário.

ASLOTEST TURBIDIMÉTIRCORef.A : 66 determinaçõesRef.B :166 determinações

PREPARO DO REAGENTE DE USOVide Instruções de uso do kit.

CONTROLE DE QUALIDADEO uso de soro controle de referência deve ser uma prática rotineira do laboratório. Recomenda-se utilizar um soro controle com valor na faixade normalidade e outro controle de valor elevado.

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS.

Page 8: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

SAC: 0800 644 6433 ww w.doles.com .br 06

AST/TGO UVManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : R-SCALIBRATION MODE : SLOPE AVGREAGENT BLANK : REAG/DILCLEANER : NOWAVELENGTH : 340nmDECIMAL POSITION : 0UNIT : UI/L

ANALYSISPOST DIL FACTOR : 2

CONC. FACTOR : NOSAMPLE CYCLE : 1

VOLUME : 16µLDILUENT NAME : H20

VOLUME : 30µLREAGENT CYCLE : 1

VOLUME : 145µL

CALCULATIONSAMPLE LIMIT : 0.700

POINT : T1REAC. DIRECTION : DECREASE

CHECK : ONCONVERS. FACTOR : 1.000

OFF SET : 0.000TEST RANGE LOW : 0

HIGH : 350NORM. RANGE LOW : * *

HIGH : * *

NUMBER OF STEPS : 1CALC. STEP A : KINSEARCHREADINGS FIRST : 3 LAST:14REACTION LIMIT : 0.2

POINT : T1CALIBRATIONCALIB. INTERVAL : * *

TIME : * *

BLANKREAG. RANGE LOW : NO

HIGH : NOBLANK RANGE LOW : -0.0020

HIGH : 0.0020

STANDARD POS : *

STD 1 :STD 2 :STD 3 :

REPLICATE : * *

DEVIATION : * *

CONTROLCS1 POS : * *

CS2 POS : * *

CS3 POS : * *

∗∗∗∗∗ Utilizar o valor assinalado do Calibrador Multiparâmetro.∗∗ ∗∗ ∗∗ ∗∗ ∗∗ Parâmetro a ser definido pelo usuário.

AST/TGO UVRef. A : 344 determinaçõesRef. B : 689 determinações

PREPARO DO REAGENTE DE USOVide Instruções de uso do kit.

CONTROLE DE QUALIDADEO uso de soro controle de referência deve ser uma prática rotineira do laboratório. Recomenda-se utilizar um soro controle com valor na faixade normalidade (soro controle N - Doles) e outro soro controle de valor elevado (soro controle P - Doles).

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS.

Page 9: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

SAC: 0800 644 6433 ww w.doles.com .br 07

CÁLCIO ARSENAZOManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : R-SCALIBRATION MODE : SLOPE AVGREAGENT BLANK : REAG/DILCLEANER : BEFOREWAVELENGTH : 600nmDECIMAL POSITION : 2UNIT : mg/dL

ANALYSISPOST DIL FACTOR : NO

CONC.FACTOR : NOSAMPLE CYCLE : 3

VOLUME : 3µLDILUENT NAME : H20

VOLUME : 20µLREAGENT CYCLE : 1

VOLUME : 300µL

CALCULATIONSAMPLE LIMIT : NOREAC. DIRECTION : INCREASE

CHECK : ONCONVERS. FACTOR : 1.000

OFF SET : 0.000TEST RANGE LOW : 0

HIGH : 20NORM. RANGE LOW : * *

HIGH : * *

NUMBER OF STEPS : 1CALC. STEP A : ENDPOINTREADINGS FIRST : 1 LAST: 6CALIBRATIONCALIB. INTERVAL : * *

TIME : * *

BLANKREAG. RANGE LOW : * *

HIGH : * *

BLANK RANGE LOW : * *

HIGH : * *

STANDARD POS : *

STD 1 :STD 2 :STD 3 :

REPLICATE : * *

DEVIATION : * *

CONTROLCS1 POS : * *

CS2 POS : * *

CS3 POS : * *

∗∗∗∗∗ Utilizar o valor assinalado do Calibrador Multiparâmetro.∗∗ ∗∗ ∗∗ ∗∗ ∗∗ Parâmetro a ser definido pelo usuário.

CÁLCIO ARSENAZO333 determinações

CONTROLE DE QUALIDADEO uso de soro controle de referência deve ser uma prática rotineira do laboratório. Recomenda-se utilizar um soro controle com valor na faixade normalidade (soro controle N - Doles) e outro soro controle de valor elevado (soro controle P - Doles).

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS.

Page 10: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

SAC: 0800 644 6433 ww w.doles.com .br 08

CKMBManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : R-SCALIBRATION MODE : FACTORREAGENT BLANK : REAG/DILCLEANER : BEFOREWAVELENGTH : 340nmDECIMAL POSITION : 0UNIT : UI/L

ANALYSISPOST DIL FACTOR : 2

CONC. FACTOR : NOSAMPLE CYCLE : 1

VOLUME : 8µLDILUENT NAME : H20

VOLUME : 10µLREAGENT CYCLE : 1

VOLUME : 197µL

CALCULATIONSAMPLE LIMIT : NOREAC. DIRECTION : INCREASE

CHECK : ONCONVERS. FACTOR : 1.000

OFF SET : 0.000TEST RANGE LOW : 0

HIGH : 500NORM. RANGE LOW : * *

HIGH : * *

NUMBER OF STEPS : 1CALC. STEP A : KINSEARCHREADINGS FIRST : 25 LAST: 37REACTION LIMIT : NO POINT : 1CALIBRATIONCALIB. INTERVAL : * *

TIME : * *

BLANKREAG. RANGE LOW : NO

HIGH : NOBLANK RANGE LOW : NO

HIGH : NO

FACTOR : 13757

CONTROLCS1 POS : * *

CS2 POS : * *

CS3 POS : * *

∗∗ ∗∗ ∗∗ ∗∗ ∗∗ Parâmetro a ser definido pelo usuário.

CKMBRef. A : 150 determinaçõesRef. B : 260 determinações

PREPARO DO REAGENTE DE USOVide Instruções de uso do kit.

CONTROLE DE QUALIDADEO uso de soro controle de referência deve ser uma prática rotineira do laboratório. Recomenda-se utilizar um soro controle com valor na faixade normalidade e outro soro controle de valor elevado.

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS.

Page 11: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

SAC: 0800 644 6433 ww w.doles.com .br 09

CK-NACManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : R-SCALIBRATION MODE : SLOPE AVGREAGENT BLANK : REAG/DILCLEANER : BEFOREWAVELENGTH : 340nmDECIMAL POSITION : 0UNIT : UI/L

ANALYSISPOST DIL FACTOR : 3

CONC. FACTOR : NOSAMPLE CYCLE : 1

VOLUME : 6µLDILUENT NAME : H20

VOLUME : 30µLREAGENT CYCLE : 1

VOLUME : 145µL

CALCULATIONSAMPLE LIMIT : 0.300

POINT : T1REAC. DIRECTION : INCREASE

CHECK : ONCONVERS. FACTOR : 1.000

OFF SET : 0.000TEST RANGE LOW : 0

HIGH : 1700NORM. RANGE LOW : * *

HIGH : * *

NUMBER OF STEPS : 1CALC. STEP A : KINSEARCHREADINGS FIRST : 9 LAST: 17REACTION LIMIT : 0.500 POINT : T1CALIBRATIONCALIB. INTERVAL : * *

TIME : * *

BLANKREAG. RANGE LOW : NO

HIGH : NOBLANK RANGE LOW : NO

HIGH : NO

STANDARD POS : *

STD 1 :STD 2 :STD 3 :

REPLICATE : * *

DEVIATION : * *

CONTROLCS1 POS : * *

CS2 POS : * *

CS3 POS : * *

∗∗∗∗∗ Utilizar o valor assinalado do Calibrador Multiparâmetro.∗∗ ∗∗ ∗∗ ∗∗ ∗∗ Parâmetro a ser definido pelo usuário.

CK NAC441 determinações

PREPARO DO REAGENTE DE USOVide Instruções de uso do kit.

CONTROLE DE QUALIDADEO uso de soro controle de referência deve ser uma prática rotineira do laboratório. Recomenda-se utilizar um soro controle com valor na faixade normalidade (soro controle N - Doles) e outro soro controle de valor elevado (soro controle P - Doles).

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS.

Page 12: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

SAC: 0800 644 6433 ww w.doles.com .br 10

CLORETO COLORIMÉTRICOManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : R-SCALIBRATION MODE : SLOPE AVGREAGENT BLANK : REAG/DILCLEANER : NOWAVELENGTH : 500nmDECIMAL POSITION : 0UNIT : mmol/L

ANALYSISPOST DIL FACTOR : 2

CONC. FACTOR : NOSAMPLE CYCLE : 1

VOLUME : 3µLDILUENT NAME : H20

VOLUME : 10µLREAGENT CYCLE : 1

VOLUME : 300µL

CALCULATIONSAMPLE LIMIT : NO

POINT : NOREAC. DIRECTION : INCREASE

CHECK : ONCONVERS. FACTOR : 1.000

OFF SET : 0.000TEST RANGE LOW : 60

HIGH : 125NORM. RANGE LOW : NO

HIGH : NONUMBER OF STEPS : 1CALC. STEP A : ENDPOINTREADINGS FIRST : CB LAST: 5

CALIBRATIONCALIB. INTERVAL : * *

TIME : * *

BLANKREAG. RANGE LOW : NO

HIGH : NOBLANK RANGE LOW : NO

HIGH : NO

STANDARD POS : *

STD 1 :STD 2 :STD 3 :

REPLICATE : * *

DEVIATION : * *

CONTROLCS1 POS : * *

CS2 POS : * *

CS3 POS : * *

∗∗∗∗∗ Utilizar o valor assinalado do Calibrador Multiparâmetro.∗∗ ∗∗ ∗∗ ∗∗ ∗∗ Parâmetro a ser definido pelo usuário.

CLORETO COLORIMÉTRICO350 determinações

CONTROLE DE QUALIDADEO uso de soro controle de referência deve ser uma prática rotineira do laboratório. Recomenda-se utilizar um soro controle com valor na faixade normalidade (soro controle N - Doles) e outro soro controle de valor elevado (soro controle P - Doles).

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS.

Page 13: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

SAC: 0800 644 6433 ww w.doles.com .br 11

COLESTEROL ENZIMÁTICO LÍQUIDOManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : R-SCALIBRATION MODE : SLOPE AVGREAGENT BLANK : REAG/DILCLEANER : BEFOREWAVELENGTH : 500nmDECIMAL POSITION : 0UNIT : mg/dL

ANALYSISPOST DIL FACTOR : 2 CONC.FACTOR : NOSAMPLE CYCLE : 1

VOLUME : 5µLDILUENT NAME : H20

VOLUME : 30µLREAGENT CYCLE : 1

VOLUME : 370µL

CALCULATIONSAMPLE LIMIT : 0.0500

POINT : T1REAC. DIRECTION : INCREASE

CHECK : ONCONVERS. FACTOR : 1.000

OFF SET : 0.000TEST RANGE LOW : 0

HIGH : 800NORM. RANGE LOW : * *

HIGH : * *

NUMBER OF STEPS : 1CALC. STEP A : ENDPOINTREADINGS FIRST : CB LAST:20

CALIBRATIONCALIB. INTERVAL : * *

TIME : * *

BLANKREAG. RANGE LOW : NO

HIGH : NO

BLANK RANGE LOW : -0.0200HIGH : 0.0200

STANDARD POS : *

STD 1 :STD 2 :STD 3 :

REPLICATE : * *

DEVIATION : * *

CONTROLCS1 POS : * *

CS2 POS : * *

CS3 POS : * *

∗∗∗∗∗ Utilizar o valor assinalado do Calibrador Multiparâmetro.∗∗ ∗∗ ∗∗ ∗∗ ∗∗ Parâmetro a ser definido pelo usuário.

COLESTEROL ENZIMÁTICO LÍQUIDORef. A : 540 determinaçõesRef. B : 1350 determinações

CONTROLE DE QUALIDADEO uso de soro controle de referência deve ser uma prática rotineira do laboratório. Recomenda-se utilizar um soro controle com valor na faixade normalidade (soro controle N - Doles) e outro soro controle de valor elevado (soro controle P - Doles).

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS.

Page 14: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

SAC: 0800 644 6433 ww w.doles.com .br 12

COLESTEROL HDLManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : R-SCALIBRATION MODE : SLOPE AVGREAGENT BLANK : REAG/DILCLEANER : BEFOREWAVELENGTH : 500nmDECIMAL POSITION : 0UNIT : mg/dL

ANALYSISPOST DIL FACTOR : 2

CONC. FACTOR : NOSAMPLE CYCLE : 1

VOLUME : 10µLDILUENT NAME : H20

VOLUME : 30µLREAGENT CYCLE : 1

VOLUME : 400µL

CALCULATIONSAMPLE LIMIT : 0.0500

POINT : T1REAC. DIRECTION : INCREASE

CHECK : ONCONVERS. FACTOR : 2.00000

OFF SET : 0.000TEST RANGE LOW : 0

HIGH : 400NORM. RANGE LOW : NO

HIGH : NONUMBER OF STEPS : 1CALC. STEP A : ENDPOINTREADINGS FIRST : T1 LAST:20

CALIBRATIONCALIB. INTERVAL : *

TIME : *

BLANKREAG. RANGE LOW : NO

HIGH : NO

BLANK RANGE LOW : -0.0200HIGH : 0.0200

STANDARD POS : *

STD 1 : 50STD 2 :STD 3 :

REPLICATE : *DEVIATION : *

CONTROLCS1 POS : *

CS2 POS : *

CS3 POS : *

∗∗∗∗∗ Parâmetro a ser definido pelo usuário.

COLESTEROL HDL40 determinações

PREPARO DO REAGENTE DE USOVide Instruções de uso do kit.

OBSERVAÇÕES:• Utilizar como calibrador o padrão que acompanha o produto Colesterol HDL - Doles.• Utilizar como amostra o sobrenadante obtido na etapa de precipitação.

Vide instruções de uso do Colesterol HDL - Doles.• Utilizar como reagente de cor o tampão enzimas do Colesterol 250 Doles ou Colesterol Enzimático Líquido Doles.

CONTROLE DE QUALIDADEO uso de soro controle de referência deve ser uma prática rotineira do laboratório. Recomenda-se utilizar um soro controle com valor na faixade normalidade (soro controle N - Doles) e outro soro controle de valor elevado (soro controle P - Doles).

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS.

Page 15: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

SAC: 0800 644 6433 ww w.doles.com .br 13

COLESTEROL HDL DIRETOManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : R-S-SR1CALIBRATION MODE : SLOPE AVGREAGENT BLANK : REAG/DILCLEANER : NOWAVELENGTH : 600nmDECIMAL POSITION : 0UNIT : mg/dL

ANALYSISPOST DIL FACTOR : NO

CONC.FACTOR : NOSAMPLE CYCLE : 1

VOLUME : 2,5µLDILUENT NAME : H20

VOLUME : 20µLREAGENT CYCLE : 1

VOLUME : 190µLSTARTR1 CYCLE : 12

VOLUME : 65µLDILUENT NAME : H20

VOLUME : 10

CALCULATIONSAMPLE LIMIT : NOREAC. DIRECTION : INCREASE

CHECK : ONCONVERS. FACTOR : 1.000

OFF SET : 0.000TEST RANGE LOW : 0

HIGH : 180NORM. RANGE LOW : * *

HIGH : * *

NUMBER OF STEPS : 1CALC. STEP A : ENDPOINTREADINGS FIRST : 11 LAST: 24

CALIBRATIONCALIB. INTERVAL : * *

TIME : * *

BLANKREAG. RANGE LOW : NO

HIGH : NOBLANK RANGE LOW : NO

HIGH : NO

STANDARD POS : * *

STD 1 : *

∗∗∗∗∗ Utilizar o valor assinalado do Calibrador que acompanha o kit.

STD 2 :STD 3 :

REPLICATE : * *

DEVIATION : * *

CONTROLCS1 POS : * *

CS2 POS : * *

CS3 POS : * *

∗∗ ∗∗ ∗∗ ∗∗ ∗∗ Parâmetro a ser definido pelo usuário.

COLESTEROL HDL DIRETO230 determinações

PREPARO DO CALIBRADORVide Instruções de uso do kit.

CONTROLE DE QUALIDADEO uso de soro controle de referência deve ser uma prática rotineirado laboratório. Recomenda-se utilizar um soro controle com valorna faixa de normalidade (soro controle N - Doles) e outro sorocontrole de valor elevado (soro controle P - Doles).

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUSPROPRIETÁRIOS.

NOTA:MAIN REAGENT : REAGENTE 1START REAGENT : REAGENTE 2

Page 16: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

SAC: 0800 644 6433 ww w.doles.com .br 14

COLESTEROL LDL DIRETOManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : R-S-SR1CALIBRATION MODE : SLOPE AVGREAGENT BLANK : REAG/DILCLEANER : NOWAVELENGTH : 550nmDECIMAL POSITION : 0UNIT : mg/dL

ANALYSISPOST DIL FACTOR : 2

CONC.FACTOR : NOSAMPLE CYCLE : 1

VOLUME : 2µLDILUENT NAME : H20

VOLUME : 20µLREAGENT CYCLE : 1

VOLUME : 200µLSTARTR1 CYCLE : 12

VOLUME : 65.0µLDILUENT NAME : H20

VOLUME : 5.0

CALCULATIONSAMPLE LIMIT : NOREAC. DIRECTION : INCREASE

CHECK : ONCONVERS. FACTOR : 1.000

OFF SET : 0.000TEST RANGE LOW : 7

HIGH : 992NORM. RANGE LOW : * *

HIGH : * *

NUMBER OF STEPS : 1CALC. STEP A : ENDPOINTREADINGS FIRST : 11 LAST: 24

CALIBRATIONCALIB. INTERVAL : * *

TIME : * *

BLANKREAG. RANGE LOW : -0.0100

HIGH : 0.1000BLANK RANGE LOW : -0.0012

HIGH : 0.0100

STANDARD POS : * *

STD 1 : *

∗∗∗∗∗ Utilizar o valor assinalado do Calibrador que acompanha o kit.

STD 2 :STD 3 :

REPLICATE : * *

DEVIATION : * *

CONTROLCS1 POS : * *

CS2 POS : * *

CS3 POS : * *

∗∗ ∗∗ ∗∗ ∗∗ ∗∗ Parâmetro a ser definido pelo usuário.

COLESTEROL LDL DIRETORef. A : 75 determinaçõesRef. B : 150 determinações

PREPARO DO CALIBRADORVide Instruções de uso do kit.

CONTROLE DE QUALIDADEO uso de soro controle de referência deve ser uma práticarotineira do laboratório. Recomenda-se utilizar um soro controlecom valor na faixa de normalidade e outro soro controle devalor elevado.

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUSPROPRIETÁRIOS.

NOTA:MAIN REAGENT : REAGENTE 1START REAGENT : REAGENTE 2

Page 17: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

CREATININA CINÉTICAManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : R-SCALIBRATION MODE : SLOPE AVGREAGENT BLANK : REAG/DILCLEANER : NOWAVELENGTH : 500nmDECIMAL POSITION : 2UNIT : mg/dL

ANALYSISPOST DIL FACTOR : 2

CONC. FACTOR : NOSAMPLE CYCLE : 1

VOLUME : 20µLDILUENT NAME : H20

VOLUME : 30µLREAGENT CYCLE : 1

VOLUME : 200µL

CALCULATIONSAMPLE LIMIT : 0.3600

POINT : T1REAC. DIRECTION : INCREASE

CHECK : ONCONVERS. FACTOR : 1.000

OFF SET : 0.000TEST RANGE LOW : 0

HIGH : 10NORM. RANGE LOW : * *

HIGH : * *

NUMBER OF STEPS : 1CALC. STEP A : ENDPOINTREADINGS FIRST : 2 LAST: 7

CALIBRATIONCALIB. INTERVAL : * *

TIME : * *

BLANKREAG. RANGE LOW : 0.0200

HIGH : 0.4000BLANK RANGE LOW : -0.0020

HIGH : 0.0040

STANDARD POS : *

STD 1 :STD 2 :STD 3 :

REPLICATE : * *

DEVIATION : * *

CONTROLCS1 POS : * *

CS2 POS : * *

CS3 POS : * *

∗∗∗∗∗ Utilizar o valor assinalado do Calibrador Multiparâmetro.∗∗ ∗∗ ∗∗ ∗∗ ∗∗ Parâmetro a ser definido pelo usuário.

CREATININA2625 determinações

PREPARO DO REAGENTE DE USOVide instruções de uso.

CONTROLE DE QUALIDADEO uso de soro controle de referência deve ser uma prática rotineira do laboratório. Recomenda-se utilizar um soro controle com valor na faixade normalidade (soro controle N - Doles) e outro soro controle de valor elevado (soro controle P - Doles).

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS.

SAC: 0800 644 6433 ww w.doles.com .br 15

Page 18: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

SAC: 0800 644 6433 ww w.doles.com .br 16

DHL-UVManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : R-SCALIBRATION MODE : SLOPE AVGREAGENT BLANK : REAG/DILCLEANER : BEFOREWAVELENGTH : 340nmDECIMAL POSITION : 0UNIT : UI/L

ANALYSISPOST DIL FACTOR : 2

CONC. FACTOR : NOSAMPLE CYCLE : 1

VOLUME : 4µLDILUENT NAME : H20

VOLUME : 10µLREAGENT CYCLE : 1

VOLUME : 200µL

CALCULATIONSAMPLE LIMIT : 0.1700

POINT : T1REAC. DIRECTION : DECREASE

CHECK : ONCONVERS. FACTOR : 1.000

OFF SET : 0.000TEST RANGE LOW : 0

HIGH : 1250NORM. RANGE LOW : * *

HIGH : * *

NUMBER OF STEPS : 1CALC. STEP A : KINSEARCHREADINGS FIRST : 3 LAST:12REACTION LIMIT : 0.1400 POINT : T1CALIBRATIONCALIB. INTERVAL : * *

TIME : * *

BLANKREAG. RANGE LOW : NO

HIGH : NOBLANK RANGE LOW : -0.020

HIGH : 0.020

STANDARD POS : *

STD 1 :STD 2 :STD 3 :

REPLICATE : * *

DEVIATION : * *

CONTROLCS1 POS : * *

CS2 POS : * *

CS3 POS : * *

∗∗∗∗∗ Utilizar o valor assinalado do Calibrador Multiparâmetro.∗∗ ∗∗ ∗∗ ∗∗ ∗∗ Parâmetro a ser definido pelo usuário.

DHL UV200 determinações

PREPARO DO REAGENTE DE USOVide Instruções de uso do kit.

CONTROLE DE QUALIDADEO uso de soro controle de referência deve ser uma prática rotineira do laboratório. Recomenda-se utilizar um soro controle com valor na faixade normalidade (soro controle N - Doles) e outro soro controle de valor elevado (soro controle P - Doles).

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS.

Page 19: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

SAC: 0800 644 6433 ww w.doles.com .br 17

FERRITINA TURBIDIMÉTRICAManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : D-R-S-SR1CALIBRATION MODE : LOGIT/LOG 5REAGENT BLANK : REAG/DILCLEANER : NOWAVELENGTH : 550nmDECIMAL POSITION : 0UNIT : µg/L

ANALYSISDILUENT NAME : H20

FACTOR : NOTIME : NOSTD : MAIN INDIRECT

MAIN STD : *POS : **

∗∗∗∗∗ Utilizar o valor assinalado do Calibrador que acompanha o kit.

FACT. STD 1 : 1.00 2 : 2.003 : 2.00 4 : 2.005 : 2.00 6 : 2.007 : NO 8 : NO

POST DIL FACTOR : 2CONC.FACTOR : NO

SAMPLE CYCLE : 1VOLUME : 28µL

DILUENT NAME : H20VOLUME : 5µL

REAGENT CYCLE : 1VOLUME : 220µL

START R1 CYCLE : 2VOLUME : 55µL

DILUENT NAME : H20VOLUME : 5µL

CALCULATIONSAMPLE LIMIT : NOREAC. DIRECTION : INCREASE

CHECK : ONANTIGEN EXCESS : NO

CONVERS. FACTOR : 1.000OFF SET : 0.000

TEST RANGE LOW : ONHIGH : ON

NORM. RANGE LOW : 20HIGH : 250

NUMBER OF STEPS : 1CALC. STEP A : KINETICREADINGS FIRST : 3 LAST : 23REACTION LIMIT : NO

CALIBRATIONCALIB. INTERVAL : * *

TIME : * *

BLANKREAG. RANGE LOW : NO

HIGH : NOBLANK RANGE LOW : NO

HIGH : NOSTANDARDS : O aparelho irá mostrar as

concentrações correspondentespara cada diluição do calibrador

REPLICATE : * *

DEVIATION : * *

CONTROLCS1 POS : * *

CS2 POS : * *

CS3 POS : * *

∗∗ ∗∗ ∗∗ ∗∗ ∗∗ Parâmetro a ser definido pelo usuário.

FERRITINA TURBIDIMÉTRICA

Ref.A : 90 determinações

Ref.B : 180 determinações

CONTROLE DE QUALIDADE

O uso de soro controle de referência deve ser uma prática

rotineira do laboratório. Recomenda-se utilizar um soro

controle com valor na faixa de normalidade e outro controle

de valor elevado.

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS

PROPRIETÁRIOS.

Page 20: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

SAC: 0800 644 6433 ww w.doles.com .br 18

FOSFATASE ALCALINA CINÉTICAManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : R-SCALIBRATION MODE : SLOPE AVGREAGENT BLANK : REAG/DILCLEANER : NOWAVELENGTH : 405nmDECIMAL POSITION : 0UNIT : UI/L

ANALYSISPOST DIL FACTOR : 3

CONC. FACTOR : NOSAMPLE CYCLE : 1

VOLUME : 4µLDILUENT NAME : H20

VOLUME : 10µLREAGENT CYCLE : 1

VOLUME : 200µL

CALCULATIONSAMPLE LIMIT : 3.000

POINT : T1REAC. DIRECTION : INCREASE

CHECK : ONCONVERS. FACTOR : 1.000

OFF SET : 0.000TEST RANGE LOW : 0

HIGH : 690NORM. RANGE LOW : * *

HIGH : * *

NUMBER OF STEPS : 1CALC. STEP A : KINSEARCHREADINGS FIRST : 3 LAST:11REACTION LIMIT : 0.3000

POINT : T1

CALIBRATIONCALIB. INTERVAL : * *

TIME : * *

BLANKREAG. RANGE LOW : NO

HIGH : NO

BLANK RANGE LOW : NOHIGH : NO

STANDARD POS : *

STD 1 :STD 2 :STD 3 :

REPLICATE : * *

DEVIATION : * *

CONTROLCS1 POS : * *

CS2 POS : * *

CS3 POS : * *

∗∗∗∗∗ Utilizar o valor assinalado do Calibrador Multiparâmetro.∗∗ ∗∗ ∗∗ ∗∗ ∗∗ Parâmetro a ser definido pelo usuário.

FOSFATASE ALCALINA CINÉTICA500 determinações

PREPARO DO REAGENTE DE USOVide Instruções de uso do kit.

CONTROLE DE QUALIDADEO uso de soro controle de referência deve ser uma prática rotineira do laboratório. Recomenda-se utilizar um soro controle com valor na faixade normalidade (soro controle N - Doles) e outro soro controle de valor elevado (soro controle P - Doles).

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS.

Page 21: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

SAC: 0800 644 6433 ww w.doles.com .br 19

FOSFATO-UVManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : R-SCALIBRATION MODE : SLOPE AVGREAGENT BLANK : REAG/DILCLEANER : BEFOREWAVELENGTH : 340nmDECIMAL POSITION : 1UNIT : mg/dL

ANALYSISPOST DIL FACTOR : NO

CONC. FACTOR : NOSAMPLE CYCLE : 1

VOLUME : 5µLDILUENT NAME : H20

VOLUME : 25µLREAGENT CYCLE : 1

VOLUME: 200µL

CALCULATIONSAMPLE LIMIT : NO

REAC. DIRECTION : INCREASECHECK : ON

CONVERS. FACTOR : 1.000OFF SET : 0.000

TEST RANGE LOW : 0HIGH : 16

NORM. RANGE LOW : * *

HIGH : * *

NUMBER OF STEPS : 1CALC. STEP A : ENDPOINTREADINGS FIRST : CB LAST: 6

CALIBRATIONCALIB. INTERVAL : * *

TIME : * *

BLANKREAG. RANGE LOW : NO

HIGH : NOBLANK RANGE LOW : 0.0050

HIGH : 0.9500

STANDARD POS : *

STD 1 :STD 2 :STD 3 :

REPLICATE : * *

DEVIATION : * *

CONTROLCS1 POS : * *

CS2 POS : * *

CS3 POS : * *

∗∗∗∗∗ Utilizar o valor assinalado do Calibrador Multiparâmetro.∗∗ ∗∗ ∗∗ ∗∗ ∗∗ Parâmetro a ser definido pelo usuário.

FOSFATO UV525 determinações

CONTROLE DE QUALIDADEO uso de soro controle de referência deve ser uma prática rotineira do laboratório. Recomenda-se utilizar um soro controle com valor na faixade normalidade (soro controle N - Doles) e outro soro controle de valor elevado (soro controle P - Doles).

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS.

Page 22: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

SAC: 0800 644 6433 ww w.doles.com .br 20

GAMA GT CINÉTICAManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : R-SCALIBRATION MODE : SLOPE AVGREAGENT BLANK : REAG/DILCLEANER : NOWAVELENGTH : 405nmDECIMAL POSITION : 0UNIT : UI/L

ANALYSISPOST DIL FACTOR : 2

CONC. FACTOR : NOSAMPLE CYCLE : 1

VOLUME : 20µLDILUENT NAME : H20

VOLUME : 30µLREAGENT CYCLE : 1

VOLUME : 200µL

CALCULATIONSAMPLE LIMIT : NOREAC. DIRECTION : INCREASE

CHECK : ONCONVERS. FACTOR : 1.000

OFF SET : 0.000TEST RANGE LOW : 0

HIGH : 300NORM. RANGE LOW : * *

HIGH : * *

NUMBER OF STEPS : 1CALC. STEP A : KINSEARCHREADINGS FIRST : 3 LAST:8REACTION LIMIT : 0.3000

CALIBRATIONCALIB. INTERVAL : * *

TIME : * *

BLANKREAG. RANGE LOW : NO

HIGH : NO

BLANK RANGE LOW : -0.020HIGH : 0.020

STANDARD POS : *

STD 1 :STD 2 :STD 3 :

REPLICATE : * *

DEVIATION : * *

CONTROLCS1 POS : * *

CS2 POS : * *

CS3 POS : * *

∗∗∗∗∗ Utilizar o valor assinalado do Calibrador Multiparâmetro.∗∗ ∗∗ ∗∗ ∗∗ ∗∗ Parâmetro a ser definido pelo usuário.

GAMA GT CINÉTICARef.A : 250 determinaçõesRef.B : 500 determinações

PREPARO DO REAGENTE DE USOVide Instruções de uso do kit.

CONTROLE DE QUALIDADEO uso de soro controle de referência deve ser uma prática rotineira do laboratório. Recomenda-se utilizar um soro controle com valor na faixade normalidade (soro controle N - Doles) e outro soro controle de valor elevado (soro controle P - Doles).

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS.

Page 23: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

SAC: 0800 644 6433 ww w.doles.com .br 21

GLICOSE ENZIMÁTICA LÍQUIDAManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : R-SCALIBRATION MODE : SLOPE AVGREAGENT BLANK : REAG/DILCLEANER : NOWAVELENGTH : 500nmDECIMAL POSITION : 0UNIT : mg/dL

ANALYSISPOST DIL FACTOR : 2

CONC. FACTOR : NOSAMPLE CYCLE : 1

VOLUME : 3µLDILUENT NAME : H20

VOLUME : 30µLREAGENT CYCLE : 1

VOLUME : 300µL

CALCULATIONSAMPLE LIMIT : 0.100

POINT : T1REAC. DIRECTION : INCREASE

CHECK : ONCONVERS. FACTOR : 1.000

OFF SET : 0.000TEST RANGE LOW : 0

HIGH : 600NORM. RANGE LOW : NO

HIGH : NONUMBER OF STEPS : 1CALC. STEP A : ENDPOINTREADINGS FIRST : T1 LAST: 8

CALIBRATIONCALIB. INTERVAL : * *

TIME : * *

BLANKREAG. RANGE LOW : 0.0000

HIGH : 0.2000BLANK RANGE LOW : -0.0080

HIGH : 0.0080STANDARD POS : *

STD 1 :STD 2 :STD 3 :

REPLICATE : **

DEVIATION : **

CONTROLCS1 POS : **

CS2 POS : **

CS3 POS : **

∗∗∗∗∗ Utilizar o valor assinalado do Calibrador Multiparâmetro.∗∗ ∗∗ ∗∗ ∗∗ ∗∗ Parâmetro a ser definido pelo usuário.

GLICOSE ENZ. LÍQUIDARef.A : 1.666 determinaçõesRef.B : 4.998 determinações

CONTROLE DE QUALIDADEO uso de soro controle de referência deve ser uma prática rotineira do laboratório. Recomenda-se utilizar um soro controle com valor na faixade normalidade (soro controle N - Doles) e outro soro controle de valor elevado (soro controle P - Doles).

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS.

Page 24: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

SAC: 0800 644 6433 ww w.doles.com .br 22

HEMOGLOBINA HbA1cManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : R-S-SR1CALIBRATION MODE : LOGIT/LOG5REAGENT BLANK : NO BLANKCLEANER : NOWAVELENGTH : 600nmDECIMAL POSITION : 2UNIT : %

ANALYSISPOST DIL FACTOR : NO

CONC.FACTOR : NOSAMPLE CYCLE : 1

VOLUME : 4µLDILUENT NAME : H20

VOLUME : 8µLREAGENT CYCLE : 1

VOLUME : 150µLSTARTR1 CYCLE : 12

VOLUME : 50.0µLDILUENT NAME : H20

VOLUME : 8.0µL

CALCULATIONSAMPLE LIMIT : NOREAC. DIRECTION : INCREASE

CHECK : ONCONVERS. FACTOR : 1.000

OFF SET : 0.000TEST RANGE LOW : ON

HIGH : ONNORM. RANGE LOW : 0.00%

HIGH : 16.00%NUMBER OF STEPS : 1CALC. STEP A : ENDPOINTREADINGS FIRST : 13 LAST: 24

CALIBRATIONCALIB. INTERVAL : * *

TIME : * *

STANDARD POS : * *

STD 1 : Valor assinalado = 0.Utilizar água ou salina.STD 2 : Valor assinalado do Calibrador HbA1c Doles Nível 01STD 3 : Valor assinalado do Calibrador HbA1c Doles Nível 02STD 4 : Valor assinalado do Calibrador HbA1c Doles Nível 03STD 5 : Valor assinalado do Calibrador HbA1c Doles Nível 04

REPLICATE : * *

DEVIATION : * *

CONTROLCS1 POS : * *

CS2 POS : * *

CS3 POS : * *

∗∗ ∗∗ ∗∗ ∗∗ ∗∗ Parâmetro a ser definido pelo usuário.

HEMOGLOBINA HbA1cRef.A : 120 determinaçõesRef.B : 200 determinações

PREPARO DO REAGENTE DE USO E CALIBRADORESVide Instruções de uso do kit.

CONTROLE DE QUALIDADEO uso de controle de referência deve ser uma prática rotineira do laboratório. Recomenda-se utilizarum controle com valor na faixa de normalidade e outro controle de valor elevado.

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS.

NOTAS:1. AS AMOSTRAS DOS PACIENTES E CALIBRADORES DEVEM SERHEMOLISADOS PREVIAMENTE CONFORME ORIENTAÇÕES DASINSTRUÇÕES DE USO.2. MAIN REAGENT : R1 LÁTEX START REAGENT : REAGENTE R2 (VIDE INSTRUÇÕES DE USO)

Page 25: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

SAC: 0800 644 6433 ww w.doles.com .br 23

MICROALBUMINÚRIA TURBIDIMÉTRICAManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : R-S-SR1CALIBRATION MODE : SLOPE AVGREAGENT BLANK : REAG/DILCLEANER : NOWAVELENGTH : 550nmDECIMAL POSITION : 0UNIT : mg/L

ANALYSISPOST DIL FACTOR : 2

CONC.FACTOR : NOSAMPLE CYCLE : 1

VOLUME : 3µLDILUENT NAME : H20

VOLUME : 10µLREAGENT CYCLE : 1

VOLUME : 280µLSTARTR1 CYCLE : 2

VOLUME : 70µLDILUENT NAME : H20

VOLUME : 20µL

CALCULATIONSAMPLE LIMIT : NOREAC. DIRECTION : INCREASE

CHECK : ONCONVERS. FACTOR : 1.000

OFF SET : 0.000TEST RANGE LOW : 0

HIGH : 80NORM. RANGE LOW : * *

HIGH : * *

NUMBER OF STEPS : 1CALC. STEP A : ENDPOINTREADINGS FIRST : 1 LAST: 6

CALIBRATIONCALIB. INTERVAL : * *

TIME : * *

BLANKREAG. RANGE LOW : NO

HIGH : NOBLANK RANGE LOW : NO

HIGH : NO

STANDARD POS : * *

STD 1 : *

∗∗∗∗∗ Utilizar o valor assinalado do Calibrador Padrão que acompanha o kit.

STD 2 :STD 3 :

REPLICATE : * *

DEVIATION : * *

CONTROLCS1 POS : * *

CS2 POS : * *

CS3 POS : * *

∗∗ ∗∗ ∗∗ ∗∗ ∗∗ Parâmetro a ser definido pelo usuário.

MICROALBUMINÚRIA TURBIDIMÉTRICA142 determinações

CONTROLE DE QUALIDADEO uso de controle de referência deve ser uma prática rotineirado laboratório. Recomenda-se utilizar um controle com valorna faixa de normalidade e outro controle de valor elevado.

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUSPROPRIETÁRIOS.

NOTA:MAIN REAGENT : DILUENTESTART REAGENT: LÁTEX

Page 26: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

SAC: 0800 644 6433 ww w.doles.com .br 24

MICROPROTE pirogalolManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : R-SCALIBRATION MODE : SLOPE AVGREAGENT BLANK : REAG/DILCLEANER : BEFOREWAVELENGTH : 600nmDECIMAL POSITION : 2UNIT : mg/dL

ANALYSISPOST DIL FACTOR : 2

CONC.FACTOR : NOSAMPLE CYCLE : 1

VOLUME : 7.0µLDILUENT NAME : H20

VOLUME : 20µLREAGENT CYCLE : 1

VOLUME : 245µL

CALCULATIONSAMPLE LIMIT : NO

POINT : T1REAC. DIRECTION : INCREASE

CHECK : ONCONVERS. FACTOR : 1.000

OFF SET : 0.000TEST RANGE LOW : 0

HIGH : 300NORM. RANGE LOW : *

HIGH : *

NUMBER OF STEPS : 1CALC. STEP A : ENDPOINTREADINGS FIRST : CB LAST: 14

CALIBRATIONCALIB. INTERVAL : *

TIME : *

BLANKREAG. RANGE LOW : NO

HIGH : NOBLANK RANGE LOW : NO

HIGH : NO

STANDARD POS : *

STD 1 : 100STD 2 :STD 3 :

REPLICATE : *

DEVIATION : *

CONTROLCS1 POS : NOCS2 POS : NOCS3 POS : NO

∗∗∗∗∗ Parâmetro a ser definido pelo usuário.

MICROPROTE pirogalolRef. A : 204 determinaçõesRef. B : 816 determinações

CALIBRAÇÃOUtilizar como calibrador, a solução padrão que acompanha o produto Microprote pirogalol.

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS.

Page 27: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

SAC: 0800 644 6433 ww w.doles.com .br 25

PCRTEST TURBIDIMÉTRICOManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : R-SCALIBRATION MODE : SLOPE AVGREAGENT BLANK : REAG/DILCLEANER : NOWAVELENGTH : 550nmDECIMAL POSITION : 1UNIT : mg/L

ANALYSISPOST DIL FACTOR : 2

CONC.FACTOR : NOSAMPLE CYCLE : 1

VOLUME : 2µLDILUENT NAME : H20

VOLUME : 10µLREAGENT CYCLE : 1

VOLUME : 300µL

CALCULATIONSAMPLE LIMIT : NOREAC. DIRECTION : INCREASE

CHECK : ONCONVERS. FACTOR : 1.000

OFF SET : 0.000TEST RANGE LOW : 0

HIGH : 75NORM. RANGE LOW : * *

HIGH : * *

NUMBER OF STEPS : 1CALC. STEP A : ENDPOINTREADINGS FIRST : 1 LAST: 6

CALIBRATIONCALIB. INTERVAL : * *

TIME : * *

BLANKREAG. RANGE LOW : NO

HIGH : NOBLANK RANGE LOW : NO

HIGH : NO

STANDARD POS : * *

STD 1 : *

∗∗∗∗∗ Utilizar o valor assinalado do Calibrador que acompanha o kit.

STD 2 :STD 3 :

REPLICATE : * *

DEVIATION : * *

CONTROLCS1 POS : * *

CS2 POS : * *

CS3 POS : * *

∗∗ ∗∗ ∗∗ ∗∗ ∗∗ Parâmetro a ser definido pelo usuário.

PCRTEST TURBIDIMÉTIRCORef.A : 66 determinaçõesRef.B :165 determinações

PREPARO DO REAGENTE DE USOVide Instruções de uso do kit.

CONTROLE DE QUALIDADEO uso de soro controle de referência deve ser uma prática rotineira do laboratório. Recomenda-se utilizar um soro controle com valor na faixade normalidade e outro controle de valor elevado.

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS.

Page 28: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

SAC: 0800 644 6433 ww w.doles.com .br 26

PCRTEST ULTRASENSÍVELManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : D-R-S-SR1CALIBRATION MODE : LOGIT/LOG4REAGENT BLANK : REAG/DILCLEANER : NOWAVELENGTH : 550nmDECIMAL POSITION : 2UNIT : mg/L

ANALYSISDILUENT NAME : H20

FACTOR : NOTIME : NOSTD : MAIN INDIRECT

MAIN STD : Conc. mg/L

POST DIL FACTOR : NOCONC.FACTOR : NO

SAMPLE CYCLE : 1VOLUME : 3,0µL

DILUENT NAME : H20VOLUME : 5,0µL

REAGENT CYCLE : 1VOLUME : 280µL

START R1 CYCLE : 18VOLUME : 20,0µL

DILUENT NAME : H20VOLUME : 0,0µL

CALCULATIONSAMPLE LIMIT : NOREAC. DIRECTION : INCREASE

CHECK : OFFANTIGEN EXCESS : NOCONVERS. FACTOR : 1.00000

OFF SET : 0.00000TEST RANGE LOW : OFF

HIGH : OFFNORM. RANGE LOW : NO

HIGH : NONUMBER OF STEPS : 1CALC. STEP A : END POINTREADINGS FIRST : 19 LAST : 35

CALIBRATIONCALIB. INTERVAL : ON REQUEST

TIME : * *

BLANKREAG. RANGE LOW : NO

HIGH : NOBLANK RANGE LOW : NO

HIGH : NO

STANDARDS POS : *

1 : C 2 : C/23 : C/4 4 : C/85 : C/16 6 : 0,07 : NO 8 : NO

REPLICATE : SINGLEDEVIATION : NO

∗∗∗∗∗ Utilizar o valor assinalado do Calibrador que acompanha o kit.

∗∗ ∗∗ ∗∗ ∗∗ ∗∗ Parâmetro a ser definido pelo usuário.

PCRTEST ULTRASENSÍVELRef.A : 120 determinaçõesRef.B : 195 determinações

PREPARO DO REAGENTE DE USOVide Instruções de uso do kit.

CONTROLE DE QUALIDADEO uso de soro controle de referência deve ser uma práticarotineira do laboratório. Recomenda-se utilizar um sorocontrole com valor na faixa de normalidade e outro controlede valor elevado.

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUSPROPRIETÁRIOS.

Page 29: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

SAC: 0800 644 6433 ww w.doles.com .br 27

PROTEÍNAS TOTAISManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : R-SCALIBRATION MODE : SLOPE AVGREAGENT BLANK : REAG/DILCLEANER : NOWAVELENGTH : 550nmDECIMAL POSITION : 1UNIT : g/dL

ANALYSISPOST DIL FACTOR : 2

CONC. FACTOR : NOSAMPLE CYCLE : 1

VOLUME : 2µLDILUENT NAME : H20

VOLUME : 10µLREAGENT CYCLE : 1

VOLUME : 250µL

CALCULATIONSAMPLE LIMIT : 0.1500

POINT : T1REAC. DIRECTION : INCREASE

CHECK : ONCONVERS. FACTOR : 1.000

OFF SET : 0.000TEST RANGE LOW : 0

HIGH : 12NORM. RANGE LOW : * *

HIGH : * *

NUMBER OF STEPS : 1CALC. STEP A : ENDPOINTREADINGS FIRST : CB LAST: 3REACTION LIMIT : NO

CALIBRATIONCALIB. INTERVAL : * *

TIME : * *

BLANKREAG. RANGE LOW : * *

HIGH : * *

BLANK RANGE LOW : * *

HIGH : * *

STANDARD POS : *

STD 1 :STD 2 :STD 3 :

REPLICATE : * *

DEVIATION : * *

CONTROLCS1 POS : * *

CS2 POS : * *

CS3 POS : * *

∗∗∗∗∗ Utilizar o valor assinalado do Calibrador Multiparâmetro.∗∗ ∗∗ ∗∗ ∗∗ ∗∗ Parâmetro a ser definido pelo usuário.

PROTEÍNAS TOTAIS2.000 determinações

PREPARO DO REAGENTE DE TRABALHOMisturar 10mL do reagente de Biureto preparado com 8 gotas de solução alcalina.

CONTROLE DE QUALIDADEO uso de soro controle de referência deve ser uma prática rotineira do laboratório. Recomenda-se utilizar um soro controle com valor na faixade normalidade (soro controle N - Doles) e outro soro controle de valor elevado (soro controle P - Doles).

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS.

Page 30: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

SAC: 0800 644 6433 ww w.doles.com .br 28

REUMATEST TURBIDIMÉTRICOManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : D-R-S-SR1CALIBRATION MODE : LOGIT/LOG 5REAGENT BLANK : REAG/DILCLEANER : NOWAVELENGTH : 600nmDECIMAL POSITION : 1UNIT : I.U/mL

ANALYSISDILUENT NAME : H20

FACTOR : NOTIME : NOSTD : MAIN INDIRECT

MAIN STD : *POS : **

∗∗∗∗∗ Utilizar o valor assinalado do Calibrador que acompanha o kit.

FACT. STD 1 : 1.00 2 : 2.003 : 2.00 4 : 2.005 : 2.00 6 : 2.007 : NO 8 : NO

POST DIL FACTOR : 2CONC.FACTOR : NO

SAMPLE CYCLE : 2VOLUME : 2µL

DILUENT NAME : H20VOLUME : 5µL

REAGENT CYCLE : 1VOLUME : 267µL

START R1 CYCLE : 1VOLUME : 27µL

DILUENT NAME : H20VOLUME : 5µL

CALCULATIONSAMPLE LIMIT : NOREAC. DIRECTION : INCREASE

CHECK : OFFANTIGEN EXCESS : NO

CONVERS. FACTOR : 1.000OFF SET : 0.000

TEST RANGE LOW : ONHIGH : ON

NORM. RANGE LOW : **HIGH : **

NUMBER OF STEPS : 1CALC. STEP A : END POINTREADINGS FIRST : 1 LAST : 7REACTION LIMIT : NO

CALIBRATIONCALIB. INTERVAL : * *

TIME : * *

BLANKREAG. RANGE LOW : NO

HIGH : NOBLANK RANGE LOW : NO

HIGH : NOSTANDARDS : O aparelho irá mostrar as

concentrações correspondentespara cada diluição do calibrador

REPLICATE : * *

DEVIATION : * *

CONTROLCS1 POS : * *

CS2 POS : * *

CS3 POS : * *

∗∗ ∗∗ ∗∗ ∗∗ ∗∗ Parâmetro a ser definido pelo usuário.

REUMATEST TURBIDIMÉTRICO

Ref. A : 66 determinações

Ref. B : 166 determinações

CONTROLE DE QUALIDADE

O uso de soro controle de referência deve ser uma prática

rotineira do laboratório. Recomenda-se utilizar um soro

controle com valor na faixa de normalidade e outro controle

de valor elevado.

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS

PROPRIETÁRIOS.

Page 31: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

SAC: 0800 644 6433 ww w.doles.com .br 29

TRIGLICÉRIDES ENZIMÁTICO LÍQUIDOManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : R-SCALIBRATION MODE : SLOPE AVGREAGENT BLANK : REAG/DILCLEANER : BEFOREWAVELENGTH : 500nmDECIMAL POSITION : 0UNIT : mg/dL

ANALYSISPOST DIL FACTOR : 2

CONC. FACTOR : NOSAMPLE CYCLE : 1

VOLUME : 4µLDILUENT NAME : H20

VOLUME : 40µLREAGENT CYCLE : 1

VOLUME : 300µL

CALCULATIONSAMPLE LIMIT : 0.0500

POINT : T1REAC. DIRECTION : INCREASE

CHECK : ONCONVERS. FACTOR : 1.000

OFF SET : 0.000TEST RANGE LOW : 0

HIGH : 800NORM. RANGE LOW : * *

HIGH : * *

NUMBER OF STEPS : 1CALC. STEP A : ENDPOINTREADINGS FIRST : CB LAST:18

CALIBRATIONCALIB. INTERVAL : * *

TIME : * *

BLANKREAG. RANGE LOW : 0.0000

HIGH : 0.1100

BLANK RANGE LOW : 0.0000HIGH : 0.1100

STANDARD POS : *

STD 1 :STD 2 :STD 3 :

REPLICATE : * *

DEVIATION : * *

CONTROLCS1 POS : * *

CS2 POS : * *

CS3 POS : * *

∗∗∗∗∗ Utilizar o valor assinalado do Calibrador Multiparâmetro.∗∗ ∗∗ ∗∗ ∗∗ ∗∗ Parâmetro a ser definido pelo usuário.

TRIGLICÉRIDES ENZIMÁTICO LÍQUIDO666 determinações

CONTROLE DE QUALIDADEO uso de soro controle de referência deve ser uma prática rotineira do laboratório. Recomenda-se utilizar um soro controle com valor na faixade normalidade (soro controle N - Doles) e outro soro controle de valor elevado (soro controle P - Doles).

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS.

Page 32: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

SAC: 0800 644 6433 ww w.doles.com .br 30

URATO ENZIMÁTICO LÍQUIDOManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : R-SCALIBRATION MODE : SLOPE AVGREAGENT BLANK : REAG/DILCLEANER : NOWAVELENGTH : 500nmDECIMAL POSITION : 1UNIT : mg/dL

ANALYSISPOST DIL FACTOR : 2

CONC. FACTOR : NOSAMPLE CYCLE : 1

VOLUME : 3.5µLDILUENT NAME : H20

VOLUME : 20µLREAGENT CYCLE : 1

VOLUME : 125µL

CALCULATIONSAMPLE LIMIT : 0.0700

POINT : T1REAC. DIRECTION : INCREASE

CHECK : ONCONVERS. FACTOR : 1.000

OFF SET : 0.000TEST RANGE LOW : 0

HIGH : 14NORM. RANGE LOW : * *

HIGH : * *

NUMBER OF STEPS : 1CALC. STEP A : ENDPOINTREADINGS FIRST : CB LAST:13

CALIBRATIONCALIB. INTERVAL : * *

TIME : * *

BLANKREAG. RANGE LOW : NO

HIGH : NOBLANK RANGE LOW : NO

HIGH : NO

STANDARD POS : *

STD 1 :STD 2 :STD 3 :

REPLICATE : * *

DEVIATION : * *

CONTROLCS1 POS : * *

CS2 POS : * *

CS3 POS : * *

∗∗∗∗∗ Utilizar o valor assinalado do Calibrador Multiparâmetro.∗∗ ∗∗ ∗∗ ∗∗ ∗∗ Parâmetro a ser definido pelo usuário.

URATO ENZIMÁTICO LÍQUIDO1280 determinações

CONTROLE DE QUALIDADEO uso de soro controle de referência deve ser uma prática rotineira do laboratório. Recomenda-se utilizar um soro controle com valor na faixade normalidade (soro controle N - Doles) e outro soro controle de valor elevado (soro controle P - Doles).

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS.

Page 33: COBAS MIRA PLUS REVISÃO: 22 (03/2013) - Doles … MIRA PLUS Rev.22 03-2013.pdf · COLESTEROL HDL ... COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS. SAC: 0800 644 6433

SAC: 0800 644 6433 ww w.doles.com .br 31

UREIA-UVManual de Automação - COBAS MIRA PLUS

GENERALMEASUMERENT MODE : ABSORBREACTION MODE : R-SCALIBRATION MODE : SLOPE AVGREAGENT BLANK : REAG/DILCLEANER : NOWAVELENGTH : 340nmDECIMAL POSITION : 0UNIT : mg/dL

ANALYSISPOST DIL FACTOR : 2

CONC. FACTOR : NOSAMPLE CYCLE : 1

VOLUME0 : 2µLDILUENT NAME : H20

VOLUME : 10µLREAGENT CYCLE : 1

VOLUME : 300µL

CALCULATIONSAMPLE LIMIT : NO

POINT :REAC. DIRECTION : DECREASE

CHECK : ONCONVERS. FACTOR : 1.000

OFF SET : 0.000TEST RANGE LOW : 0

HIGH : 300NORM. RANGE LOW : NO

HIGH : NONUMBER OF STEPS : 1CALC. STEP A : KINETICREADINGS FIRST : 2 LAST:4REACTION LIMIT : 0.1600

POINT : T1

CALIBRATIONCALIB. INTERVAL : * *

TIME : * *

BLANKREAG. RANGE LOW : NO

HIGH : NOBLANK RANGE LOW : -0.040

HIGH : 0.0020

STANDARD POS : *

STD 1 :STD 2 :STD 3 :

REPLICATE : * *

DEVIATION : * *

CONTROLCS1 POS : * *

CS2 POS : * *

CS3 POS : * *

∗∗∗∗∗ Utilizar o valor assinalado do Calibrador Multiparâmetro.∗∗ ∗∗ ∗∗ ∗∗ ∗∗ Parâmetro a ser definido pelo usuário.

UREIA UV666 determinações

PREPARO DO REAGENTE DE USOVide Instruções de uso do kit.

CONTROLE DE QUALIDADEO uso de soro controle de referência deve ser uma prática rotineira do laboratório. Recomenda-se utilizar um soro controle com valor na faixade normalidade (soro controle N - Doles) e outro soro controle de valor elevado (soro controle P - Doles).

COBAS MIRA PLUS É MARCA REGISTRADA DE SEUS PROPRIETÁRIOS.