    SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0

    5-2I2

    5.3 SYSTEM SETTING PROCEDURES

    5.3.1 GENERAL

    This paragraph provides controller system setting procedure.

    NOTICE: Please follow by the procedures of item 5.3.3. if it is unnecessary, please do not

    change data of system setting to avoid hazards.

    5.3.2 BRIEF INTRODUCTION

    Controller system setting procedures listed below:

    1. LED Lightness set

    2. Air temperature sensor compensation

    3. Air flow temperature sensor compensation

    4. Skin temperature sensor compensation

    5. Machine Number

    6. Motor speed alarm value

    7. Set value of Low deviation for air temperature

    8. Set value of High deviation for air temperature

    9. Set value of Low deviation for skin temperature

    10. Set value of High deviation for skin temperature

    11. Minimum set value of air temperature

    12. Maximum set value of air temperature

    13. Minimum set value of skin temperature

    14. Maximum set value of skin temperature

    15. Set value of Over temperature alarm (37 )

    16. Set value of Over temperature alarm (37 )

    17. Set value of air flow Over temperature alarm (37 )

    18. Set value of air flow Over temperature alarm(37 )

  • 7/31/2019 YP-90 Series Ajuste

    2/22

    SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0

    5.3.3 PROCEDURE

    At first, connected power supply, press and hold Set key for 2s as soon as switch on until Air

    temprature indicator displays --.-, and no display on Skin and Set temprature indicator. And then,

    press Up or Down until Air temprature indicator displays 78.6, after then, press Set key to enter into

    the system set screen, or press Silence/Reset key to exit.

    After entering into the set screen, Set temprature indicator displays set code, Air temprature

    indicator displays the specific set data.

    Turn on the power switch

    5-3I2

    Press and hold the for 3s

    Enter password

    Press key, EXIT Press or key to enter password 78.6,

    Then press key to confirm

    Error Password check

    YES

    System setting screen

    Press key or Key, set temperature indicator displays

    the set code

    Press or Key can change

    current set value

    END

    Press key Press and together

    SAVE and EXIT Return to original set

    Brightness set of indicator

    Figure 5.2

  • 7/31/2019 YP-90 Series Ajuste

    3/22

    SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0

    5-4I2

    CODE 00.1 LED Lightness set

    Indicate the lightness , and set range is 00.1~01.5. The larger the lighter.

    CODE 00.2Air temperature sensor compensation

    Set value of temperature compensation for air temperature sensor , and set range is -5.0~05.0 ,

    the first figure of set value means the increasing and decreasing value. For Example, 00.2 means

    that the air temperature adds 0.2 as compensation. -1.3 means that the air temperature reduces

    1.3 as compensation.

    CODE 00.3Air flow temperature sensor compensation

    temperature compensation for air flow temperature sensor , and set range is -5.0~05.0 , the

    first figure of set value means the increasing and decreasing value. For Example, 00.2 means that

    the air flow temperature adds 0.2 as compensation. -1.3 means that the air flow temperature

    reduces 1.3 as compensation.

    CODE 00.4 Skin temperature sensor compensation

    Set value of temperature compensation for skin temperature sensor , and set range is

    -5.0~05.0 , the first figure of set value means the increasing and decreasing value. For Example,

    00.2 means that the skin temperature adds 0.2 as compensation. -1.3 means that the skin

    temperature reduces 1.3 as compensation.

    CODE 00.5 Machine Number

    Set method of machine number, and set range is 00.0~02.0. For example, 00.1 measn No. 1

    machine, 01.5 means No. 15.

    CODE 00.6 Motor speed alarm valueSet method of fan speed, and set range is 00.0~25.5. Every 0.1 means 20 rpm. For example,

    05.0 means 1000 rpm. Alarm will be active if the fan speed is lower than this value.

    CODE 00.7 Set value of Low deviation for air temperature

    Set method of low deviation alarm in air mode, and set range is 00.0~05.0. For example, 00.8

    means 0.8 .

    CODE 00.8 Set value of High deviation for air temperature

    Set method of high deviation alarm in air mode, and set range is 00.0~05.0. For example, 00.8

    means 0.8 .

    CODE 00.9 Set value of low deviation for skin temperature

    Set method of low deviation alarm in skin mode, and set range is 00.0~05.0. For example, 00.8

    means 0.8 .

    CODE 01.0 Set value of high deviation for skin temperature

    Set method of high deviation alarm in skin mode, and set range is 00.0~05.0. For example, 00.8

    means 0.8 .

  • 7/31/2019 YP-90 Series Ajuste

    4/22

    SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0

    5-5I2

    CODE 01.1 Minimum set value of air temperature

    Set method of Minimum set value of air temperature, and set range is 20.0~34.0. For example,

    25.0 means Minimum set value 25.0 .

    CODE 01. 2 Maximum set value of air temperature

    Set method of Maximum set value of air temperature in mode of>37.0 , and set range is

    37.0~39.0. For example, 39.0 means Maximum set value 39.0 .

    CODE 01.3 Minimum set value of skin temperature

    Set method of Minimum set value of skin temperature, and set range is 30.0~34.0. For example,

    34.0 means Minimum set value 25.0 .

    CODE 01.4 Maximum set value of skin temperature

    Set method of Maximum set value of skin temperature in mode of>37.0 , and set range is

    37.0~39.0. For example, 39.0 means Maximum set value 39.0 .

    CODE 01.5 Set value of Over temperature alarm A (37

    Set method of Over temperature alarm in air mode (set temperatuer37, and set range is

    37.5~39.0. For example, 38.0 means over-temp alarm value 38.0 .

    CODE 01.6 Set value of Over temperature alarm B (37)

    Set method of Over temperature alarm in air mode (set temperatuer37, and set range is

    38.5~40.0. For example, 40.0 means over-temp alarm value 40.0 .

    CODE 01.7 Set value of air flow Over temperature alarm A (37

    Set method of alarm value for air flow blocked (set temperatuer 37, and set range is

    38.5~48.0. For example, 45.0 means alarm value for air flow blocked 45.0

    CODE 01.8 Set value of air flow Over temperature alarm B (37)

    Set method of alarm value for air flow blocked (set temperatuer >37, and set range is 40.5~52.0.

    For example, 45.0 means alarm value for air flow blocked 45.0

  • 7/31/2019 YP-90 Series Ajuste

    5/22

    SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0

    5.4 TROUBLESHOOTING PROCEDURES

    5.4.1 GENERAL

    Troubleshooting guides for the Controller are provided in Paragraph 5.4.3 and Paragraph 5.4.4.

    Paragraph 5.4.3 provides the alarm code when the controller alarms. Paragraph 5.4.4 provides

    troubleshooting in the form of flowcharts.

    It is confirmed that the system settings are correct before troubleshooting.

    5.4.2 TEST EQUIPMENT REQUIRED

    The test equipment listed below is required for troubleshooting the Controller. Equivalent test

    equipment may be substituted.

    Digital VOM, FLUKE 8060A

    31.0 Air Temperature Simulation

    5-6I2

    Resistor

    Resistor R=23166/0.25W/0.5%

    31.0 Air flow Temperature Simulation

    Resistor

    Resistor R=23166/0.25W/0.5%

    31.0 skin Temperature Simulation

    Resistor R=23166/0.25W/0.5%

    5.4.3 TROUBLESHOOTING

    All alarms have the relevant alarm code except for the power ailure alarm, please refer to Table 5.1.

    TABLE 5.1 ERROR CODES

    ALARM CODE TYPEEXPLANATION AND

    CORRECTIVE ACTION

    E0.1 Air temperature sensor failure alarm Refer to Flow chart 5.1

    E0.2 Isolated temperature sensor alarm Refer to Flow chart 5.2

    E0.3 Air temperature deviation failure alarm Refer to Flow chart 5.3

    Resistor

  • 7/31/2019 YP-90 Series Ajuste

    6/22

    SERVICE MANUAL FOR I FANT INCUBATOR EDITION/REVISION B/0N

    5-7I2

    TABLE 5.1 ERROR CODES (Continued)

    ALARM CODE TYPEEXPLANATION AND

    CORRECTIVE ACTION

    E0.4 Air flow sensor alarm Refer to Flow chart 5.4

    E0.5 Skin temperature sensor failure alarm Refer to Flow chart 5.5

    E0.6 Over temperature failure alarm Refer to Flow chart 5.6

    E0.7 Air circulation failure alarm Refer to Flow chart 5.7

    E0.8 Wrong position of skin sensor alarm Check the position of skin sensor

    E0.9 Fan motor stop alarm Refer to Flow chart 5.8

    E1.0 High deviation temperature failure alarmCheck whether there is heat sourcenearby, if yes, please leave far awayfrom the heat source.

    E1.1 Low deviation temperature failure alarmCheck the close condition of all doors

    and panels

    H0.1 ROM failure alarm inside of Main MCU Replace Main MCU

    H0.2Internal system failure alarm inside ofMain MCU

    Replace Main MCU

    H0.3 communication failure alarm

    1. Check the connection of circuitbetween the main MUC and auxiliaryMCU2. Replace main MUC and auxiliaryMCU

    H0.4 ROM failure alarm inside auxiliary MCU Replace auxiliary MCU

    H0.5

    Internal system failure alarm inside

    auxiliary MCU Replace auxiliary MCU

    H0.6 E2PROM failure alarm Replace Main MCU

    H0.7 SRAM failure alarm Replace U13 on mian board

    H0.8 Main ADC failure alarm Refer to Flow chart 5.9

    H0.9 Auxiliary ADC failure alarm Refer to Flow chart 5.10

    H1.0 Heat system failure alarm Refer to Flow chart 5.11

    H1.1 Real timer RTC failure alarm Replace U20 on mian board

    H1.2 Keyboard fault alarm When the key board of the control panelon controller short-circuit alarms, pleasereplace the control panel.

    H1.3 Battery failure alarm

    Inner storage battery failure alarm:1. Check the wire connection betweenthe inner storage battery and the powerboard.2. Replace the inner storage battery.

    Power failure alarm Refer to Flow chart 5.12

  • 7/31/2019 YP-90 Series Ajuste

    7/22

    SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0

    5.4.4 TROUBLESHOOTING FLOW CHART

    Before troubleshooting, forst, please switch off, and disconnect the power cable, and then open the

    top cover of warmer module and the back cover of column. And then connect the power cable and turn

    onthe switch, at the same time, make sure that power +5V, +12V of controller, element, and all cables

    are in good condition.In working state, press Up Key and Air mode Key , you can find the isolated

    temperature and air flow temperarture in the air and set temperature indicator respectively. So it is easy

    for service man to find the cause of trouble.

    AIR TEMPERATURE SENSOR FAILURE ALARM E0.1

    no

    Good conection? Check connection of air temperature sensor,

    and the wiring of all circuit board

    The 31.00.2 Air Temperature Simulation (decade

    Equal to 23166182)is instead of the air

    temperature sensor, turn on power supply

    the temperature display is

    31.0 and add the no

    compensation temperature alarm E0.1? Replace the air temperature sensor

    of system set procedure

    yes

    on the main board U11-6 is 2.3V? no check the connect to R21, U1-6

    Yes

    CHECK U11

    FLOW CHART 5.1 AIR TEMPERATURE SENSOR FAILURE

    5-8I2

  • 7/31/2019 YP-90 Series Ajuste

    8/22

    SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0

    ISOLATED TEMPERATURE SENSOR FAILURE ALARM E0.2

    ``

    no

    Good conection? Check connection of isolated temperature

    sensor, and the wiring of all circuit board

    The 31.00.2 Air Temperature Simulation (decade

    Equal to 23166

    182

    )is instead of the isolatedtemperature sensor, turn on power supply

    See 5.4.4, the isolated no

    temperature is 31.0 alarm E0.2? Replace the isolated temperature sensor

    yes

    no

    on the main board U21-6 is 2.3V? check the connect to R16, U21-6

    Yes

    Check U21

    FLOW CHART 5.2 ISOLATED TEMPERATURE SENSOR FAILURE

    5-9I2

  • 7/31/2019 YP-90 Series Ajuste

    9/22

    SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0

    Air temperature deviation failure alarm E0.3

    ``

    noGood conection? Check connection of isolated temperature

    sensor, and the wiring of all circuit board

    The 31.00.2 Air Temperature Simulation (decade

    Equal to 23166182)is instead of the air and isolated

    temperature sensor, turn on power supply

    See 5.4.4, check whether

    The air temeprature and no

    isolated tempertire is alarm E0.3? Replace sensor

    added the temperature

    compensation based on 31.0

    yes

    U11-6, no

    on the main board U21-6 is 2.3V? check the connect to R16 and U21-6,

    R21 and U11-6

    Yes

    Replace U11 or U21

    FLOW CHART 5.3 AIR TEMPERATURE DEVIATION FAILURE

    5-10I2

  • 7/31/2019 YP-90 Series Ajuste

    10/22

    SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0

    Air flow sensor failure alarm E0.4

    noGood conection? Check connection of sensor,and the wiring of all

    circuit board

    The 31.00.2 Air Temperature Simulation (decade

    Equal to 23166182)is instead of air flow

    temperature sensor, turn on power supply

    See 5.4.4, check whether

    The air temeprature and no

    isolated tempertire is alarm E0.4? Replace air flow sensor

    added the temperature

    compensation based on 31.0

    yes

    no

    on the main board U11-8 is 2.3V? check the connect to R23 and U11-8

    Replace U11

    FLOW CHART 5.4 AIR FLOW SENSOR FAILURE

    5-11I2

  • 7/31/2019 YP-90 Series Ajuste

    11/22

    SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0

    SKIN TEMPERATURE SENSOR FAILURE ALARM E0.5

    Check connection between skin sensor

    Good conection? Socket and circuit board, the connection Of all

    circuit board

    The 31.00.2 Air Temperature Simulation (decade

    Equal to 23166182)is instead of air flow

    temperature sensor, turn on power supply

    the temperature display no

    is 31.0, added alarm E0.5? Replace the skin temperature sensor

    the temperature compensation

    in system set procedure

    yes

    no

    on the main board U11-7 is 2.3V? check connect Relay R22 to U1-7

    Yes

    Check U11

    FLOW CHART 5.5 SKIN TEMPERATURE SENSOR FAILURE

    5-12I2

  • 7/31/2019 YP-90 Series Ajuste

    12/22

    SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0

    Over-temp failure E0.6

    yes

    Is there heat source nearby? Far away from the heat source

    5-13I2

    no

    yes

    Air and isolated temperature Check air and isolated temperature sensor

    Sensor is OK?

    no

    Replace air and isolated temperature sensor

    FLOW CHART 5.6 OVER TEMPERATURE FAILURE

    Check whether the inlet

    Or outlet is blocked

    Air circulation failure alarm E0.7

    Check the air flow condition of inlet and outlet

    FLOW CHART 5.7 AIR CIRCULAATION FAILURE

  • 7/31/2019 YP-90 Series Ajuste

    13/22

    SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0

    Fan Motor Failure Alarm E0.9

    Motor rotating no Motor input terminal no Check the power cord

    Is AC power?

    yes

    yes

    on the power Pulse at no Check the Capacitor no

    supply board U5-2? fan detected CM is good Replace the capacitor CM

    board

    yes yes

    5-14I2

    no Replace the Motor

    on the power Pulse at Check U5

    supply board U5-3

    yes

    on the main no

    board Pulse U12-16? check the connection between the main board

    and the power supply board

    yes

    Check U12

    FLOW CHART 5.8 FAN MOTOR FAILURE ALARM( SHEET 1 OF 2)

  • 7/31/2019 YP-90 Series Ajuste

    14/22

    SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0

    Slow Motor Speed Alarm E0.9

    yes

    Motor input no

    Voltage is ok? Check the power input

    yes

    no

    Capacitor Replace the capacitor CM

    CM is good

    yes

    Replace the motor

    FLOW CHART 5.8 FAN MOTOR FAILURE ALARM ( SHEET 2 OF 2)

    5-15I2

  • 7/31/2019 YP-90 Series Ajuste

    15/22

    SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0

    Connection of U11

    and U12 is OK?

    Check the jumper wire of

    U11-13, R5, R6

    noPin of U11-13 is

    +2.5V?

    no

    Check R5, R6Crosspoint of R5

    and R6 is +2.5V?

    Main ADC failure alarm H0.8

    On main board

    On main board

    Check the connecting wireOn main board

    Replace U11

    FLOW CHART 5.9 MAIN ADC FAILURE ALARM

    5-16I2

  • 7/31/2019 YP-90 Series Ajuste

    16/22

    SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0

    5-17I2

    Check the connecting wireConnection of U21and U18 is OK?

    Check the jumper wire of

    U21-9, R14, R15

    noPin of U21-9 is

    +2.5V?

    no

    Check R14, R15Crosspoint of R14

    and R15 is +2.5V?

    Auxiliary ADC failure alarm H0.9

    On main board

    On main board

    On main board

    Replace U21

    FLOW CHART 5.10 AUXILIARY ADC FAILURE ALARM

  • 7/31/2019 YP-90 Series Ajuste

    17/22

    SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0

    Heat circuit failure alarm H1.0

    Disconnet power supply, measure resistance of twoterminals for heater

    5-18I2

    yes

    Replace heater

    no

    Heater is open circuit or

    short circuit?

    Conneat power board and main board

    Element number from here is on

    power board

    Connect pin 1 (ssr-ctr) and pin 3

    (relay-ctr) on J8 to VCC power make two

    conrol signals high level

    no

    Base of triode Q2, Q3

    is >0.7V?

    Check triode Q2, Q3

    and resistor R8,R9

    yes

    nothere +12V on solid

    relay and control

    terminal of relay K1

    Replace triode

    Q2 or Q3

    yes

    To sheet 2

    FLOW CHART 5.11 HEAT SYSTEM FAILURE (SHEET 1 OF 4)

  • 7/31/2019 YP-90 Series Ajuste

    18/22

    SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0

    From sheet 1

    Output terminal of solid

    relay and relay K1 can

    work?

    no Replace solid relay

    or relay K1

    yes

    Voltage of U1-2 is

    > 5mV

    yes

    no

    Check inductor T1

    Base of triode Q1

    >0V?

    yes

    noBase of triode Q1

    >0.7V?

    yes

    no

    Voltage of U1-1

    is >2V?Check U1,R1 ,C4

    Check R2,R6,C5

    no

    Repalce triode Q1

    Connet pin 1 (ssr-ctr) and pin3 (relay-ctr) on J8 to GND

    make two control signals high level

    yes

    To sheet 3

    FLOW CHART 5.11 HEAT SYSTEM FAILURE (SHEET 2 OF 4)

    5-19I2

  • 7/31/2019 YP-90 Series Ajuste

    19/22

    SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0

    From sheet 2

    Base of triode Q1,Q3

    is 0V?

    yes

    Check triode Q2,Q3

    and resistor R8, R9

    no

    Base of triode Q1

    is 5V?

    yes

    noBase of triode Q1

    is 0V?

    Yes

    no

    Voltage U1-1 is

    0V?

    Yes

    no

    Voltage U1-2 is

    0V?

    no

    yesIs there +12V on

    control terminal of solid

    relay and relay K1

    Replace triode

    Q2 or Q3

    Check inductor T1

    Check U1,R1,C4

    Check R2, R6, C5

    no

    Replace triode Q1

    yes

    To sheet 4

    FLOW CHART 5.11 HEAT SYSTEM FAILURE (SHEET 3 OF 4)

    5-20I2

  • 7/31/2019 YP-90 Series Ajuste

    20/22

    SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0

    From sheet 3

    The pins on porer

    board J8-1, 3, 4 and

    J2-1, 3, 4 on main

    board works?

    Switch off during measuring

    no

    Replace connecting wire

    yes

    Connect power board and main board

    Connect pins U8-2, 11 to VCC power

    Element number from here is onmainboard

    Yes

    Pin of U8-9, 18 is

    5V?

    no

    Check U8

    Measure the voltage of U15-2 for self-checking

    yes

    no

    Replace U15

    yes yes

    Replace main board

    Still alarming after

    replacing U15

    Voltage is changingfrom low level to highlevel?

    FLOW CHART 5.11 HEAT SYSTEM FAILURE (SHEET 4 OF 4)

    5-21I2

  • 7/31/2019 YP-90 Series Ajuste

    21/22

    SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0

    Power Failure Alarm

    5-22I2

    Is there AC power on

    input socket?Check outside power input

    No

    Yse

    Initial trans for mer is

    AC ower?

    No Check the connecting wire of

    trans for mer and ower cord,

    and Fuse F1, F2, F3 on AC

    entry board

    Yes

    No

    Pin 3 on U2, U3 is +5V?

    No

    Cross point of R17,R18 is +2.5V?

    No

    Check U2, U3 and prefix circuit

    Yes

    On power board

    Pin4 of U4 is

    +12V?Check U4 and prefix circuitOn power board

    Yes

    Check R17, R18

    Yes

    Secondary output oftransformer is 9V, 15V? Replace transformer

    On power board

    Yes

    To sheet 2

    FLOW CHART 5.12 POWER FAILURE ALARM (SHEET 1 OF 2)

  • 7/31/2019 YP-90 Series Ajuste

    22/22

    SERVICE MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0

    From sheet 1

    NoCrosspoint of

    R19, R20 is 2.5V?On power board Replace R19, R20

    Yes

    Crosspoint of

    R21, R22 is

    2.1V?

    No

    Replace R21, R22On power board

    5-23I2

    Yes

    Pins7, 1 of U6

    is high level?

    No

    On power board Replace U6

    Yes

    NoPin5 of U8 is low

    level?Check the connection of power board

    and main board

    On main board

    Yes

    Pin4 of U8 is

    low level?

    No

    Replace U8On main board

    Yes

    Replace main board

    FLOW CHART 5.12 POWER FAILURE ALARM (SHEET 2 OF 2)